OCTAVIUS Detector 1600 SRS Type of product: Two-dimensional array with 1521 liquid-filled ionization chambers Application: IMRT and VMAT patient plan verification Machine-specific QA Online beam adjustment Measuring quantities: Absorbed dose rate and absorbed dose Range of use: 0.1 - 24 Gy/min Resolution: 0.1 mGy, 0.1 mGy/min Dead time: Zero Display cycle: 100 - 800 ms Type of detectors: Plane-parallel, liquid-filled ionization chambers Detector layout: Center area (6.5 cm x 6.5 cm): spacing 2.5 mm Outer area (15 cm x 15 cm): spacing 5 mm Nominal response: 16 nC/Gy Size of detectors: 2.5 mm x 2.5 mm x 0.5 mm (0.003 cm³) Active detector area: 15 cm x 15 cm Outer dimensions: 300 mm x 420 mm x 22 mm(w x d x h)重量:5.9 kg